Home · Why single-use? The evidence

why single-use speculum evidence

Why single-use? The evidence

In short. Metal instruments can remain contaminated after cleaning, and reprocessing is not always done correctly, even in advanced systems. No opened source documents a confirmed infection outbreak transmitted by a classic vaginal speculum. Single-use is not the only solution.

Gexa Scientific Team · Published 2026-11-24 · Last updated 2026-11-24

Metal instruments can still carry material after cleaning, and reprocessing is not always carried out correctly, including in advanced health systems. That sentence is about contamination and about lapses. It is not a confirmed outbreak transmitted by a vaginal speculum, and single-use is not the only solution.

The paragraphs below follow the opened sources. Each one says what the source measured, and what it did not. Analogous evidence, from other devices, is labeled as analogous.

Limits that stay on this page

No source opened for this page documents a confirmed patient-to-patient infection transmitted by a classic reusable vaginal speculum. The large exposure investigations say the opposite of an outbreak claim: NHS Lanarkshire reported no known cases transmitted in this way; Toronto Public Health confirmed no bloodborne disease from the Park clinic among 622 results; the Milne abstract reported negative screens in 17 interviews; the Hong Kong decision does not mention an infection; the VA report called the risk negligible.

Estes 2007 also found that autoclaving left HPV DNA undetectable in all 70 samples. A sentence about “metal failing” that does not separate correct sterilization from a lapse is not accurate.

DNA is not infection. Ferenczy, Gallay, Casalegno, M'Zali and Leroy measured contamination. Gallay and Ferenczy say infectiousness was not shown.

Transvaginal probes are not vaginal specula. Saudi surgical-site rates are not evidence about a named device. No verified vaginal-speculum outbreak was found for Saudi Arabia, the Gulf, other Arab countries, Africa, or Latin America in the pages that were opened. Absence of a published report is not proof that nothing has ever happened. It does forbid a claim of “more than twenty speculum outbreaks.”

Single-use is not the only solution. A reusable instrument reprocessed as intended is a different case from a lapse.

Saudi Arabia and the GCC

CSSD auditing, GDIPC annual reports 2022 and 2023

Context only. These figures measure compliance with reprocessing standards. They are not infection counts and are not specific to vaginal specula.

According to the 2022 annual report of the General Directorate of Infection Prevention and Control (GDIPC), Saudi Ministry of Health, CSSD auditing visits generated 752 reports from 414 healthcare facilities, with an overall general compliance rate of 83.72%. According to the 2023 GDIPC annual report, overall CSSD compliance was 83.60%, ranging from 91% in Al-Qurayyat and Jeddah to 72% in Al-Madinah and Al-Baha.

Source record

National surgical-site infection surveillance, Saudi Arabia, 2022

The source does not attribute these events to vaginal specula or to a named reprocessing failure. They must not be presented as evidence of harm from reusable specula.

According to the National Healthcare-Associated Infections Report 2022 (PubMed PMID 37741012) and the SSI table in the 2022 GDIPC annual report, the surgical-site infection rate was 0.87 per 100 surgeries among 86,324 monitored operations (751 events), including 640 events after 71,919 caesarean sections (0.89 per 100).

Source record

United States, Europe, Canada, and Hong Kong

Autoclave sterilization of instruments used on women with cervical neoplasia

Direct. Vaginal specula and biopsy instruments. DNA, not a patient infection.

According to Estes, Kirby and Huh (2007, Journal of Lower Genital Tract Disease; PubMed PMID 17194945), HPV DNA was detected on 51.1% (23/45) of used vaginal specula and 55% (11/20) of used biopsy instruments. All 70 autoclaved samples were negative. One instrument processed with heated glass beads and a Cidex bath, and presumed sterile, was positive for HPV-16 DNA. The study measured residual DNA, not patient infection, and also found autoclaving reduced HPV DNA to undetectable levels in this assay.

Source record

HPV DNA on gynecologic instruments and fomites

Direct evidence about biopsy forceps and cryoprobe tips, not a vaginal speculum. The authors state that whether HPV on fomites is infectious was not evaluated. DNA is not a patient infection.

According to the PubMed abstract of Ferenczy et al. (PMID 2555753, Obstetrics & Gynecology, 1989), HPV DNA was found on 37% (23/62) of biopsy forceps before, and 1.6% (1/62) after, 30 minutes in 30% tincture of Savlon; and on 23% (5/22) of cryoprobe tips before, and 4.5% (1/22) after, 1 minute of 90% ethanol.

Source record

IPAC investigation, community gynecology clinic

Direct. A gynecology clinic lapse. No attributed bloodborne infection in 622 reported results.

According to Toronto Public Health’s official IPAC final report on the community gynecology clinic of Dr Esther Park, an IPAC lapse was identified, including no reprocessing quality-assurance records on site. As of 31 August 2025, approximately 2,600 letters had been sent to patients who underwent higher-risk procedures between 10 October 2020 and 10 October 2024. TPH received results for 622 patients: no cases of bloodborne diseases were confirmed to have been the result of exposure at the clinic. TPH described the public-health risk as very low. Pap smears or vaginal examination with a speculum were listed among lower-risk procedures. The owner began replacing reusable instruments with single-use devices where possible. That is a corrective step, not a measured drop in infections.

Source record

IPAC lapse, transvaginal ultrasound probes, Dr Nicholas VJ Pairaudeau

Direct evidence about transvaginal probes, which are not vaginal specula. The posted reports do not give a count of attributed infections.

According to Toronto Public Health’s official initial and interim reports on the clinic of Dr Nicholas VJ Pairaudeau (2026), the lapse included failure to properly clean and disinfect transvaginal ultrasound probes, plus sterilization quality-assurance deficiencies. The clinic was ordered to stop procedures requiring multi-use instruments processed in the current sterilizer, and to stop transvaginal ultrasound.

Source record

Patient safety issues, endovaginal transducers, Mayagüez

Direct for endovaginal probes, which are not specula. No attributed infections listed.

According to VA Office of Inspector General report 09-03055-103 (16 March 2010), endovaginal transducers at the Mayaguez outpatient clinic were not submitted to high-level disinfection after each patient for approximately two years, and the same condition existed at the San Juan hospital. The risk to patients was determined to be negligible. The report does not list confirmed infections attributed to this lapse.

Source record

High-risk HPV DNA on endovaginal ultrasound probes after low-level disinfection

Direct evidence about ultrasound probes, not specula. The study documents residual DNA, not documented patient-to-patient infection.

According to Casalegno et al. (PLOS ONE, 2012, Lyon University Hospital), after endovaginal ultrasound and the then-recommended low-level disinfection, 6 samples (3.0%) contained high-risk HPV types. Among pre-use samples, HPV was detected in 6 (2.7%), including 4 (1.9%) with high-risk types.

Source record

Microbial contamination of transvaginal probes after low-level disinfection

Direct evidence about probes, not specula. The paper does not document a patient infection outbreak.

According to M'Zali et al. (PLOS ONE, 2014), after low-level disinfection of covered transvaginal probes, HPV DNA was found on 13% of samples (7% in nuclease-resistant form), and Staphylococcus aureus was cultured from 4%.

Source record

Infectious risk of endovaginal and rectal ultrasound probes

Probe evidence, not specula. The pooled 3.1% infection figure after transrectal prostate biopsy is specific to that procedure and is not a vaginal-speculum rate. Very few cases had an established contamination route.

According to Leroy’s systematic review and meta-analysis (Journal of Hospital Infection, 2013), the pooled prevalence of pathogenic bacteria on endovaginal or rectal probes after low-level disinfection was 12.9% (95% CI 1.7–24.3), and 1.0% (0.0–10.0) for HPV, HSV or CMV.

Source record

Human papillomavirus (HPV) contamination of gynaecological equipment

Direct. Surfaces and one reusable speculum. DNA is not shown to be infectious virus.

According to Gallay et al. (Sexually Transmitted Infections, 2015, Switzerland), 32 of 179 samples (18%) from gynecological equipment and surfaces were HPV-positive despite routine cleaning. The colposcope had the highest contamination risk. One reusable speculum was HPV-positive; the authors could not exclude sampling contamination or a laboratory error. They state there is no evidence that contaminated surfaces carry infectious viruses. They also suggested that disposable specula could be preferred to help prevent contamination. The paper did not measure a reduction in patient infections after a switch.

Source record

Reusable metal specula not always sterilised, Clarkston Medical Centre

He described a theoretical risk that is very low, and said there are no known cases of infection transmitted in this way. This is not a confirmed speculum outbreak.

According to Dr John Logan, Consultant in Public Health Medicine for NHS Lanarkshire, as reported by BBC News on 18 February 2004, reusable metal vaginal specula were not always sterilised before being used again between April 2000 and November 2003 at Clarkston Medical Centre, Airdrie, and officials were trying to contact 260 women.

Source record

Women’s experience of a look-back after inadequate speculum decontamination

The full paper was not opened, so no figures beyond the abstract are used. The abstract refers to look-back when estimated risks to health are low. It does not document a confirmed infection.

According to the 2006 abstract by Paul Milne that was opened, after inadequate decontamination of vaginal specula was identified at a GP surgery, over 400 women were offered screening for chlamydia, hepatitis B and hepatitis C. All 17 women interviewed had negative screening results.

Source record

Inquiry decision NC/441/7/B, unsterile steel speculum used once

The decision documents a patient-safety breach. It does not report that an infection occurred. The centre’s routine was a disposable plastic speculum first.

According to the Nursing Council of Hong Kong decision of 5 February 2018 (case NC/441/7/B), on 29 September 2016 a reusable stainless-steel vaginal speculum that had been cleaned and packed to be sterilized, and that lacked a sterilization label, was used on one patient for cervical screening.

Source record

Analogous

Infections associated with reprocessed duodenoscopes

Analogous evidence only. Duodenoscopes are not vaginal specula. Do not present this as a speculum outbreak.

According to the FDA page on infections associated with reprocessed duodenoscopes, if these devices are not thoroughly cleaned and disinfected, tissue or fluid from one patient can remain, and in rare cases this can lead to patient-to-patient transmission of infection. Newer-model sampling found 1.1% positivity for high-concern organisms, better than the 4 to 6% observed with similar older models.

Source record

Sustainable practice: Switching to reusable vaginal speculums

A limit, not a selling point. No iatrogenic-infection reports found by these authors for reusable instruments.

According to Drayton, Smith and Ratnappuli (BMJ, 2023), disinfection and sterilisation are effective methods of microbial inactivation, and the authors wrote that no reports in the literature describe iatrogenic infection from reusable medical instruments. That sentence is a limit on any claim about a correctly reprocessed speculum. It is broader than the literature on other devices, such as duodenoscopes and laparoscopic instruments, and it is kept here as a limit, not waved aside.

Source record

India and Africa

These records are analogous from the first line. They are not studies of the vaginal speculum.

Mycobacterium chelonae port-site infections after chemical disinfection of laparoscopic instruments

Analogous evidence only. This is a laparoscopic-instrument outbreak, not a vaginal-speculum outbreak.

According to Vijayaraghavan et al. (Journal of Hospital Infection; PubMed 17046106), at a hospital in Bangalore, India, 145 laparoscopy port-site infections due to Mycobacterium chelonae were found in 35 patients over six weeks. The source was rinse water used after chemical disinfection.

Source record

Audit of reusable surgical-device processing, CHU Yalgado Ouédraogo

Analogous evidence only. The source does not report attributed infections and does not mention vaginal specula.

According to Josephine et al. (2013, Antimicrobial Resistance and Infection Control; PMC3688477), an audit of reusable surgical-device processing at CHU Yalgado Ouédraogo, Burkina Faso, found cleaning in 4.65% of observed cases and high rates of non-compliant practice, in a hospital without a recommended central sterilization department.

Source record

What is known after a move toward single-use

No opened primary study measured fewer patient infections after metal vaginal specula were replaced with single-use plastic specula. What follows is a recommendation, a clinic practice, or a corrective step, or it is analogous evidence from other devices. It is not the only solution.

Gallay and colleagues suggested disposable specula might help prevent contamination, and they also said there is no evidence that contaminated surfaces carry infectious virus. Toronto Public Health recorded that the Park clinic owner replaced reusable instruments with single-use devices where possible after a lapse. The Hong Kong Nursing Council decision records that the centre’s routine was a disposable plastic speculum for a cervical smear, with steel only when clinically indicated. None of these sources measured an infection-rate reduction.

The WHO aide-memoire says a device marked single-use should not be reused, and it classes a reusable vaginal speculum as intermediate risk. That is guidance, not a trial of infection rates.

Analogous evidence, other devices

These results are not vaginal-speculum results.

Aide-memoire: decontamination and reprocessing of medical devices

This is reprocessing guidance, not a trial showing that disposable specula reduce patient infections.

According to the WHO aide-memoire on decontamination and reprocessing of medical devices (WHO/UHL/IHS/IPC/2022.4, citing the 2016 manual), single-use devices should only be used as recommended by the manufacturer and should not be reused. The same document lists reusable vaginal specula as intermediate-risk items requiring at least high-level disinfection.

Source record

Disposable versus reusable cystoscopes in an outpatient practice

Analogous evidence only. The study is retrospective and non-randomized, and it concerns cystoscopy, not vaginal specula.

According to the 2023 Urology Practice abstract (PubMed PMID 37228224), conversion to disposable cystoscopes in an outpatient practice was associated with fewer 30-day post-procedure encounters (11/494, 2.2% versus 35/506, 7.1%) and fewer positive urine cultures (1/494, 0.2% versus 17/506, 3.4%) than reusable cystoscopes.

Source record

Cross-contamination and infection with reusable flexible bronchoscopes

Analogous bronchoscope evidence, not vaginal-speculum evidence. The authors’ zero-risk line is an assumption, not a measured speculum result.

According to Mouritsen et al. (Anaesthesia, 2020; PMC7079200), 16 studies of reusable flexible bronchoscopes (2,351 patients; 3,120 procedures) yielded a 2.8% incidence of cross-contamination or infection (476 contamination events; 86 linked infections). The authors expected the single-use risk to be 0% because no cross-contamination cases had been reported, and they caution that a zero-risk assumption is limited.

Source record

References

  1. General Directorate of Infection Prevention and Control (GDIPC), Saudi Ministry of Health (2023). CSSD auditing, GDIPC annual reports 2022 and 2023. GDIPC, Ministry of Health, Saudi Arabia. https://psj1-ifn-s3-ifn01.jed-s3.bluvalt.com/files/Feb2024/1-The%20Indicator%20and%20Reports%20Unit/Reports/GDIPC%20ANNUAL%20REPORT%202022.pdf
  2. GDIPC, Saudi Ministry of Health; National Healthcare-Associated Infections Report (2022). National surgical-site infection surveillance, Saudi Arabia, 2022. PubMed PMID 37741012; GDIPC annual report 2022. https://pubmed.ncbi.nlm.nih.gov/37741012/
  3. Estes JM, Kirby TO, Huh WK (2007). Autoclave sterilization of instruments used on women with cervical neoplasia. Journal of Lower Genital Tract Disease. 10.1097/01.lgt.0000229565.19364.cc
  4. Ferenczy A, Bergeron C, Richart RM (1989). HPV DNA on gynecologic instruments and fomites. Obstetrics & Gynecology. https://pubmed.ncbi.nlm.nih.gov/2555753/
  5. Toronto Public Health (2025). IPAC investigation, community gynecology clinic. City of Toronto. https://www.toronto.ca/community-people/health-wellness-care/health-inspections-monitoring/infection-prevention-and-control-investigations-ipac/
  6. Toronto Public Health (2026). IPAC lapse, transvaginal ultrasound probes, Dr Nicholas VJ Pairaudeau. City of Toronto. https://www.toronto.ca/community-people/health-wellness-care/health-inspections-monitoring/infection-prevention-and-control-investigations-ipac/
  7. VA Office of Inspector General (2010). Patient safety issues, endovaginal transducers, Mayagüez. VA OIG report 09-03055-103. https://www.vaoig.gov/sites/default/files/reports/2010-03/VAOIG-09-03055-103.pdf
  8. Casalegno J-S and colleagues (2012). High-risk HPV DNA on endovaginal ultrasound probes after low-level disinfection. PLOS ONE. 10.1371/journal.pone.0048137
  9. M'Zali F and colleagues (2014). Microbial contamination of transvaginal probes after low-level disinfection. PLOS ONE. 10.1371/journal.pone.0093368
  10. Leroy S (2013). Infectious risk of endovaginal and rectal ultrasound probes. Journal of Hospital Infection. https://pubmed.ncbi.nlm.nih.gov/23201397/
  11. Gallay C, Miranda E, Schaefer S, Catarino R, Jacot-Guillarmod M, Menoud PA, Guerry F, Achtari C, Sahli R, Vassilakos P, Petignat P (2016). Human papillomavirus (HPV) contamination of gynaecological equipment. Sexually Transmitted Infections. 10.1136/sextrans-2014-051977
  12. NHS Lanarkshire, Dr John Logan, reported by BBC News (2004). Reusable metal specula not always sterilised, Clarkston Medical Centre. BBC News, 18 February 2004. https://news.bbc.co.uk/2/hi/uk_news/scotland/3499909.stm
  13. Paul Milne (2006). Women’s experience of a look-back after inadequate speculum decontamination. Quality in Primary Care. https://www.primescholars.com/articles/womens-experience-of-a-lookback-exercise-following-inadequate-decontamination-of-vaginal-specula.pdf
  14. Nursing Council of Hong Kong (2018). Inquiry decision NC/441/7/B, unsterile steel speculum used once. Nursing Council of Hong Kong. https://www.nchk.org.hk/filemanager/en/pdf/decision_of_inquiry_for_Chan_Lai_wa_case_441_website.pdf
  15. U.S. Food and Drug Administration (2022). Infections associated with reprocessed duodenoscopes. FDA. https://www.fda.gov/medical-devices/reprocessing-reusable-medical-devices/infections-associated-reprocessed-duodenoscopes
  16. BMJ (2023). Sustainable practice: Switching to reusable vaginal speculums. BMJ. 10.1136/bmj-2023-075779
  17. Vijayaraghavan R, Chandrashekhar R, Sujatha Y, Belagavi CS (2006). Mycobacterium chelonae port-site infections after chemical disinfection of laparoscopic instruments. Journal of Hospital Infection. https://pubmed.ncbi.nlm.nih.gov/17046106/
  18. Josephine J, Ndoye B, Guissou IP, Zongo FS, Dehainsala H (2013). Audit of reusable surgical-device processing, CHU Yalgado Ouédraogo. Antimicrobial Resistance & Infection Control. https://pmc.ncbi.nlm.nih.gov/articles/PMC3688477/
  19. World Health Organization (WHO) (2022). Aide-memoire: decontamination and reprocessing of medical devices. WHO/UHL/IHS/IPC/2022.4. https://iris.who.int/server/api/core/bitstreams/d9c93a0a-4014-41a1-a8e8-3bc6fdd2ce40/content
  20. Urology Practice, 2023 (authors not printed on the opened abstract) (2023). Disposable versus reusable cystoscopes in an outpatient practice. Urology Practice. 10.1097/upj.0000000000000410
  21. Mouritsen JM, Ehlers L, Kovaleva J, Ahmad I, El-Boghdadly K (2020). Cross-contamination and infection with reusable flexible bronchoscopes. Anaesthesia. https://pmc.ncbi.nlm.nih.gov/articles/PMC7079200/

In this cluster